TotalAlert Infinity™ NFPA Medical Gas Notification System
Home Button Information Button Alarm Button (Red)
Alarm Configuration History Button Figure 46: Fault Status
Button
Button
Figure 45: Icon Toolbar NOTE:
Icons on the toolbar: The Normal toolbar will automatically
reappear when all faults are resolved.
1. Home button. This icon button toggles to
the MAIN screen.
Users can touch any flashing red light to go to
2. Alarm button. This icon button toggles to the ACTIVE ALARM screen.
the ACTIVE ALARM screen.
3. Configuration button. This icon button
toggles to the CONFIGURATION screen.
4. Information button. This icon button
toggles to the INFORMATION screen.
5. History button. This icon button toggles
to the HISTORY screen.
The button that corresponds to the screen
currently in use will be blue.
4.1.6 Fault Status
If a gas sensor is measuring above or below
the alarm level, the high or low pressure in-
dicator on the gas badge will flash for the gas
pressure out of range.
If a gas sensor becomes disconnected or
shorted, both high and low indicators for
that gas will flash and an error code and mes-
sage will be displayed.
If a configured source signal becomes an
open circuit, the source badge and the signal
status will change from green to blinking red.
The Icon toolbar will appear, and the Alarm
button will be flashing red. By default the
screen will change to the Active Alarm
Screen.
4-3 4107 9016 58.02