TotalAlert Infinity™ NFPA Medical Gas Notification System



                                                                 4.3 Setting Up an Area Alarm














               Figure 57: Normal Status Message screen
            2.  Touch the Erase button on the keyboard
               to delete the existing message.
            3.  Enter the new word or message.
            4.  Touch the Save button.
                                                                       Figure 59: Area Alarm User Interface
            Panel ID Description

            To change the description of the location of         4.3.1 Configuring Gas Badges
            the alarm panel:
                                                                                        ™
                                                                 The TotalAlert Infinity  area alarm is designed
            1.  Touch the [B] button that is located to the     to allow configuration of eight gas badges us-
               right of the numeric display (Figure 56).        ing the B60 board.
            2.  Touch the Erase button on the keyboard
               to delete the existing description.              To configure the first badge to custom set-
            3.  Enter the new description.                      tings, touch the [1] button in the numeric dis-
            4.  Touch the Save button.                          play of the configuration screen (Figure 56).





















                 Figure 58: Panel ID Description screen

















                                                             4-7                                4107 9016 58.02