TotalAlert Infinity™ NFPA Medical Gas Notification System



           Primary Settings Tab                                Initialize Modules Tab

           The PRIMARY SETTINGS tab allows you to              The PRIMARY SETTINGS screen will now
           select the alarm type, initialize boards, and       display with ALARM TYPE checked, INITIALIZE
           select gas inputs.                                  MODULES active, and the other tabs grayed
           •  Touch the PRIMARY SETTINGS tab                   out.





           Alarm Type Tab

           The ALARM TYPE screen defines which type
           of alarm will be displayed and drives later
           setup functions.







                                                                       Figure 75: Initialize Modules Tab

                                                               •  Touch the INITIALIZE MODULES tab.
                                                               The INITIALIZE MODULES screen is used to set
                                                               up the boards that are installed. The boards
                                                               function as the input/output hardware of the
                                                               alarm.
                      Figure 73: Alarm Type Tab
           •  Touch the ALARM TYPE tab.
           •  Choose a code style based on code re-
              quirements.




                                                     Save
                                                     Erase




                                                                      Figure 76: Initialize Modules Screen
                                                               At this point, you have two options:
             Figure 74: Choosing Code and Alarm Styles
                                                               1.  Either touch the Scan button to prompt
           •  Touch the AREA tab                                   the alarm to auto fill the screen for you.
           •  Touch the Save button when you see               OR
              checks beside your selections to save the        2.  Complete the screen by selecting the
              settings and return to the PRIMARY SET-              boards installed in the alarm.
              TINGS screen.

           Note: To clear your selections, touch the
           Erase button.
           4107 9016 58.02                                 4-12