TotalAlert Infinity™ NFPA Medical Gas Notification System
• Touch the GAS SENSORS tab.
Figure 77: B60 board selected
Figure 79: Gas Input Screen
Note:
THE GAS INPUT / B60 board must be select- Users may select up to eight different gases
ed during the configuration of an area alarm. on the GAS INPUT screen. These gases cor-
respond to the digital gas sensors wired into
• Touch the Save button when you see the GAS INPUT / B60 board (see section
checks beside your selections. 2.4.2).
• To select the first gas, touch the D01 =
Gas Sensors Tab NONE/BLANK tab.
The PRIMARY SETTINGS screen will now dis-
play with ALARM TYPE and INITIALIZE MOD-
ULES checked, GAS SENSORS active, and the
last tab grayed out.
Figure 80: Gas Sensors Screen
Figure 78: Gas Sensors Tab