TotalAlert Infinity™ NFPA Medical Gas Notification System
4.4.2 Initial Setup of an Unconfig- • Enter the security password. (The default
ured Master Alarm password is 121212.)
• Touch the Enter button.
The SETUP INFORMATION screen is the first • Touch Continue to advance to the
screen that users will see on an unconfigured CONFIGURATION screen.
alarm.
(The MAIN screen is the first screen that users Note:
will see on a configured alarm). If you would like to change the password,
• Touch the Configuration button. touch Change Password. Enter a 4-10 digit
password, and touch the Enter button. Con-
firm the new password. Touch the X button
in the upper right to close the SECURITY
PASSWORD screen.
Configuration
Button
Figure 92: Setup Information Screen
Entering the Password
The ENTER SECURITY PASSWORD screen ap- Figure 94: Changing the Security Password
pears.
• The CONFIGURATION screen will display
as unconfigured and ready for setup
Primary Settings Tab
The PRIMARY SETTINGS tab allows you to
select the alarm type and initialize boards.
• Touch the PRIMARY SETTINGS tab.
Figure 93: Entering the Security Password
4107 9016 58.02 4-18