TotalAlert Infinity™ NFPA Medical Gas Notification System



            Each source badge has room for eight source          B2 = Signal Settings Tab
            signal inputs (B-J) (Figure 88). Two tabs are        Signal Text
            used for the detailed configuration of each
            source signal input. The two tabs for each           •  Touch the NONE/BLANK tab
            source signal input are paired and named
            alphabetically (e.g., B1 and B2 / C1 and C2 /
            D1 and D2 / etc.).
            Users can scroll to the first tab for each source
            signal input by using the right and left arrows
            at the bottom of the screen. To jump quickly
            to the second and third screens, simply touch
            a letter button on the grid on the left that
            appears on that screen (e.g., the D button
            jumps to the second screen and the G button
            jumps to the third screen).
            The first tab for each source signal input is                     Figure 91: Signal Text
            highlighted and the second grayed out to             To select source signal input text:
            ensure that users complete the setup in the
            correct order. The second tab in each pair           •  Scroll through the options using the
            will become available to you as you advance             right and left arrows at the bottom of the
            through setup.                                          screen.
                                                                 •  Touch the tab for the appropriate option,
            B1 = Alarm Point Tab                                    or select a blank to create a custom alarm.
                                                                 •  Touch the Save button when you see a
                                                                    check beside your selection to close and
                                                                    return to the SIGNAL SETTINGS screen.
                                                                 •  Touch the X button in the upper right
                                                                    to close and return to the CONFIGURE
                                                                    (SOURCE) screen.

                                                                  Active Alarm Instruction:

                                                                  To set facility instructions for active alarms,
                                                                  touch either the GENERAL MESSAGE or OP-
                                                                  TION MSG tab. The alarm defaults to GEN-
                                                                  ERAL MESSAGE. See Section 4.6 for how to
                      Figure 90: Alarm Points Tab
                                                                  set up custom instructions.
            To identify the location where the wires are
            landed:                                              Configuring Additional Source Badges

            1.  Touch the tab for the appropriate board          Configure alarm points on additional source
               and terminal block. Ex: B10 Board, X01            badges by touching the number on the
               Terminal Block                                    numeric display that corresponds with the

            2.  Touch the Save button when you see a             source badge and repeating the steps listed
               check beside your selection to go auto-           above.
               matically to the SIGNAL SETTINGS screen.