TotalAlert Infinity™ NFPA Medical Gas Notification System
Alarm Type Tab
The ALARM TYPE screen defines which type
of alarm will be displayed and drives later
setup functions.
Figure 97: Initialize Modules Screen
• Touch the INITIALIZE MODULES tab.
The INITIALIZE MODULES screen is used to set
up the boards that are installed. The boards
Figure 95: Alarm Type Tab function as the input/output hardware of the
alarm.
• Touch the ALARM TYPE tab.
• Choose a code style based on code re-
quirements.
• Touch the MASTER tab.
Save
Erase
Figure 98: Initialize Modules Screen
At this point, you have two options:
1. Either touch the Scan button to prompt
Figure 96: Choosing Code and Alarm Styles the alarm to autofill the screen for you.
• Touch the Save button when you see OR
checks beside your selections to close the 2. Complete the screen by selecting the
screen and return to the PRIMARY SET- boards installed in the alarm.
TINGS screen.
Touch the Save button when you see checks
Note: To clear your selections, touch the beside your selections.
Erase button.
The PRIMARY SETTINGS screen will now dis-
Initialize Modules Tab play with ALARM TYPE and INITIALIZE MOD-
The PRIMARY SETTINGS screen will now dis- ULES highlighted and checked and the other
play with ALARM TYPE checked and INITIAL- tabs grayed out.
IZE MODULES active. • Touch the X button in the upper right to
close and return to the CONFIGURATION
screen.
4-19 4107 9016 58.02