TotalAlert Infinity™ NFPA Medical Gas Notification System



            Alarm Type Tab

            The ALARM TYPE screen defines which type
            of alarm will be displayed and drives later
            setup functions.











                                                                       Figure 97: Initialize Modules Screen
                                                                •  Touch the INITIALIZE MODULES tab.

                                                                The INITIALIZE MODULES screen is used to set
                                                                up the boards that are installed. The boards

                       Figure 95: Alarm Type Tab                function as the input/output hardware of the
                                                                alarm.
            •  Touch the ALARM TYPE tab.
            •  Choose a code style based on code re-
               quirements.
            •  Touch the MASTER tab.





                                                       Save

                                                       Erase

                                                                       Figure 98: Initialize Modules Screen
                                                                At this point, you have two options:

                                                                1.  Either touch the Scan button to prompt
              Figure 96: Choosing Code and Alarm Styles             the alarm to autofill the screen for you.
            •  Touch the Save button when you see               OR
               checks beside your selections to close the       2.  Complete the screen by selecting the
               screen and return to the PRIMARY SET-                boards installed in the alarm.
               TINGS screen.
                                                                Touch the Save button when you see checks
            Note: To clear your selections, touch the           beside your selections.
            Erase button.
                                                                The PRIMARY SETTINGS screen will now dis-
            Initialize Modules Tab                              play with ALARM TYPE and INITIALIZE MOD-

            The PRIMARY SETTINGS screen will now dis-           ULES highlighted and checked and the other
            play with ALARM TYPE checked and INITIAL-           tabs grayed out.
            IZE MODULES active.                                 •  Touch the X button in the upper right to
                                                                    close and return to the CONFIGURATION
                                                                    screen.


                                                             4-19                               4107 9016 58.02