TotalAlert Infinity™ NFPA Medical Gas Notification System
4.4 Setting Up a Master Alarm Note:
The master alarm monitors whether source Maximum wired input signals allowed is 60.
equipment is working properly. Two indepen-
dent hard wired alarms are required per facil- To configure alarm points in the first badge,
ity to meet the 24-hour monitoring require- touch the [1] button in the numeric display
ment in the NFPA 99 standard. (Figure 83).
The information for the signals connected in
the field needs to be entered into the alarm
using the touch screen interface.
Figure 84: Configuration Source Screen
A = Gas Color Code Tab
Figure 82: Master Alarm User Interface
To set up or change the configuration of an
alarm, follow the steps outlined under Alarm
Type Tab.
4.4.1 Configuring Source Badges
™
The TotalAlert Infinity master alarm is de-
signed to allow the setup and monitoring of
12 source badges. Forty source locations with Figure 85: Gas Color Code Tab
up to 8 source signal inputs for each location
can be configured (maximum of 20 locations To select the color code for the appropriate
per gas type). gas based on your standard and country:
1. Touch the A = GAS COLOR CODE tab
(Figure 84).
2. Scroll through the gas types using the
right and left arrows at the bottom of the
screen.
3. Touch the tab for the appropriate code, or
select a “custom” tab to create a unique
source badge.
4. Touch the Save button when you see a
Figure 83: Configuration Screen check beside your selection.
4-15 4107 9016 58.02