TotalAlert Infinity™ NFPA Medical Gas Notification System



                                                                H = Zero Offset Tab
           Active Alarm Instruction:
           To set facility instructions for active alarms,
           touch either the GENERAL MESSAGE or OP-
           TION MSG tab. The alarm defaults to GEN-
           ERAL MESSAGE. See Section 4.6 for how to
           set up custom instructions.


           G = High Alarm Settings Tab







                                                                           Figure 68: Zero Offset Tab

                                                               In some cases, facility dial gauges may show
                                                               slightly different readings than the TotalAlert
                                                               Infinity™ alarm (e.g., 46 PSI versus 44 PSI). To
                                                               match the TotalAlert Infinity™ alarm exactly to
                                                               a facility dial gauge:

                                                               1.  Select the offset amount.
                 Figure 67: High Alarm Settings Tab            2.  Touch the Save button to close the screen.

           You can remove the high alarm set point or          Configuring Additional Gas Badges
           customize the high alarm set point.
                                                               Configure additional gas badges by touching
           To remove the high alarm set point:
                                                               the number on the main configuration screen
           1.  Touch the NONE tab.                             that corresponds with the gas badge and
           2.  Touch the check on the CONFIRM SET-             repeating the steps listed above.
              TINGS CHANGE screen.
           To customize the high alarm set point:

           1.  Touch the ADJUST tab.
           2.  Enter the new set point.
           3.  Touch the Save button.