TotalAlert Infinity™ NFPA Medical Gas Notification System



           WIFI Tab                                            4.2 Alarm Configuration

           Touch the Wireless Ethernet (WIFI) tab to
           view WIFI Status (On or Off), SSID identifi-        To set up or change the configuration of an
           cation, and IP Address. WIFI Status can be          alarm, start on the CONFIGURATION screen.
           turned On or Off by touching the designated         Users are always prompted to enter the pass-
           tabs.                                               word when accessing this screen.


                                                                                                            Close
                                                                                                            Clear all
                                                                                                            entered
                                                                                                            text








                                                                           Backspace      Enter
                                                                Figure 55: Password entry on Configuration screen
                         Figure 53: WIFI Tab
                                                               Type the default password (121212), and
           History Screen                                      touch the Enter button. You will be prompt-

           To view up to 100 alarm events on screen,           ed to change the password if you have not
           touch the history button located on the icon        done so previously.
           toolbar. User can easily scroll to view the         The CONFIGURATION screen displays.
           events by touching up/down arrows in the            Normal Status Message/Panel ID
           bottom right corner of the screen.                  Description

                                                               The TotalAlert Infinity™ alarm allows for cus-
                                                               tomization of certain labels and messages.
                                                               Normal Status Message

                                                               To change the message that is communicated
                                                               when the alarm is operating in normal status:

                                                               1.  Touch the [A] button that is located to the
                                                                   left of the numeric display (Figure 56).





                       Figure 54: History Screen











                                                                        Figure 56:  Configuration Screen


           4107 9016 58.02                                  4-6