TotalAlert Infinity™ NFPA Medical Gas Notification System



            2.4.2 Locally Installed Sensors                      2.4.3 Remotely Installed Sensors in
                                                                 Pipeline
            1.  Remove sensor module from shipping
               carton.                                           1.  Remove sensor from shipping carton.
            2.  Connect sensor with DISS fitting to the
               appropriate DISS check at the top of the          2.  Connect sensor with DISS fitting to the
                                                                    appropriate DISS Check on the hospital
               rough-in box. Verify Gas ID labels match
               between the sensor and piping to ensure              pipeline or BeaconMedæs Zone Valve box.
                                                                    Verify Gas ID labels match between the
               no cross connections occur. Repeat this              sensor and piping to ensure no cross con-
               process for all sensors within alarm panel.
            3.  Connect wires into input terminals of B60           nections occur.
               board (see section 2.3.5).                        3.  Wire nut pigtail to field installed wiring.
                                                                    Note polarity of wiring and corresponding
                                                                    field wiring color or number for later (see
                                                                    section 2.3.5).


                                                                  NOTE:
                                                                  Do not ground shield drain wire at sensor.
















                   Figure 33:  Local Sensor Mounting

             NOTE:
                                                                  Figure 34:  Remote Sensor Module Connection
             If gas sensor wires are landed on the terminal
             blocks in the same order as defined by the
             model number of the alarm panel, then initial
             setup will be easier and gas sensor channels
             will not need to be reconfigured.”  Example
             T3-A10-OAV means “O” Oxygen is defaulted to
             D01, “A” Medical Air is defaulted to D02, etc.
             See Section 1.6.