TotalAlert Infinity™ NFPA Medical Gas Notification System



           2.4 Finish Assembly


           2.4.1 Front Panel Mounting






                           ATTENTION:

              Observe Precautions for Handling
            ELECTROSTATIC SENSITIVE DEVICES





           1.  Remove front panel assembly from pack-
              aging.

           2.  Remove front panel mounting screws
              from alarm panel rough-in box assembly                               Ground Lug
              (Figure 31).                                                Figure 31: Front Panel Screws

           3.  Remove lanyard mounting nut from alarm
              panel back box assembly (Figure 31).
           4.  Attach front panel to alarm panel rough-
              in box using screws removed in step 2
              (Figure 32).
           5.  Attach lanyard to alarm panel rough-in
              box using screws removed in step 3 (Fig-
              ure 32).
           6.  Remove nut from front panel grounding
              lug.
           7.  Install front panel grounding wire using
              nut removed in step 6.
           8.  Connect the grey cable on the front panel
              to the open power/communication port
              on the power supply.                                Figure 32:  Attach Front Panel and Lanyard


            NOTE:
            Harness connector is keyed to prevent
            incorrect orientation. However, use care to
            ensure correct pin alignment.