TotalAlert Infinity™ NFPA Medical Gas Notification System



            2.3.12 Field Wiring Cable Shield                     2.3.13 BACnet
            Grounding
                                                                 The BeaconMedæs TotalAlert Infinity™ NFPA
            All field wiring cable shields must be ground-       control systems utilize the TCP/ IP family of
            ed inside alarm panel back box.                      protocols for communication and each alarm
                                                                 has a unique IP address.
            Holes are provided in the top and bottom of          Each BACnet enabled TotalAlert Infinity™
            the alarm rough-in box for grounding screws          alarm conforms to the BACnet protocol.
            (Figure 30).
                                                                 •  Protocol Implementation Conformance
            Shields from several cables may be wrapped              Statement
            together and crimped into one screw (by oth-         Each BACnet enabled TotalAlert Infinity™
            ers).                                                alarm has available a Protocol Implementa-
                                                                 tion Conformance Statement (PICS). The PICS
                                                                 is a written document created by Beacon-
                                                                 Medæs that identifies the particular options
                                                                 specified by BACnet that are implemented in
                                                                 the TotalAlert Infinity™ alarm.
                                                                 •  BACnet Conformance

                                                                 BACnet Vendor ID: 543
                                                                 Alarm Profile: BACnet Smart Sensor (B-SS)
                                                                 BIBB (BACnet Interoperability Building
                                                                 Blocks)
                                                                 Supported:
                         Grounding Holes Provided
                                                                 •  Read Property - B (DS-RP-B)
                Figure 30:  Cable Shield Grounding-Area
                                                                 •  Dynamic Device Binding - B (DM-DDB-B)

                                                                 •  Dynamic Object Binding - B (DM-DOB-B)
                 CAUTION:
                                                                 Object Type Support: Analog value, Binary
            Keep shield wires to ground screw as short           value, Device.
            as possible so they can not touch front              Data Line Layer Options: ISO 8802-3 (“Ether-
            panel circuit boards when front panel is             net”) LAN
            closed.
                                                                 Character Set Support: ISO 10646 (UTF-8)
                                                                 Encoding