Page 51 - Small Animal Internal Medicine, 6th Edition
P. 51

CHAPTER 2   Diagnostic Tests for the Cardiovascular System   23



                              Long-axis view 1                   for ventricular measurements; however, it may be more dif-
                                                                 ficult to verify that the beam is transecting the LV at its
  VetBooks.ir                                 RVO  PV            maximum dimension. Precise positioning of the ultrasound
                                                                 beam within the heart (perpendicular to the structures to
                                               RC
                                          LV
                                                                 for accurate  M-mode  measurements  and calculations. For
                                               NC    AO          be measured) and clear endocardial images are essential
                                               LA                example, papillary muscles within the LV must be avoided
                                                                 when measuring free-wall thickness.  Fig. 2.10 illustrates
                                                                 standard M-mode views. In cases where the M-mode cursor
                                                                 cannot be optimally aligned (e.g., in animals with focal or
                              Long-axis view 2                   asymmetric hypertrophy), wall thickness measurements
                                                                 from  2-D  images  are preferred, although  in animals with
                                                                 a high heart rate, obtaining true end-diastolic frames can
                                                                 be uncertain.
                                                  RV
                                            LV
                                                                 Common Measurements and
                                      C a         RA             Normal Values
                                        V C             RAu
                                                                 The standard dimensions measured with M-mode and their
                                                                 timing also are indicated in  Fig. 2.10. The “leading edge”
                                                                 technique is used when possible (i.e., from the edge closest
                                                                 to the transducer [leading edge] of one side of the dimension
                              Long-axis view 3                   to the leading edge of the other). In this way, only one endo-
                                                                 cardial thickness is included in the measurement. LV wall
                                              RVO                and interventricular septal thicknesses, as well as LV chamber
                                                  PV             dimensions, should be determined at the level of the chordae
                                           LV      PA            tendineae rather than the apex or mitral valve level. Mea-
                                                                 surements also may be taken from 2-D images if they are of
                                                LA               high resolution, and frames from the appropriate times in
                                                                 the cardiac cycle are used. Body size greatly influences echo-
                                                                 cardiographic measurements, especially in dogs. However,
                                                                 the relationship between body weight or surface area and
                                                                 cardiac dimensions is not linear. Rather, cardiac linear
            FIG 2.9
            Left cranial parasternal long-axis views optimized for aortic   dimensions are more closely related to body length, which is
                                                                                                        ⅓
            root (above), right atrium and auricle (middle), and right   proportional to body weight to the ⅓ power (BW ). Allome-
            ventricular outflow and main pulmonary artery (below).   tric scaling has been used to generate guidelines for common
            These views are used to evaluate the heart base and can   cardiac dimension measurements in normal dogs (see Table
            provide good Doppler signals for tricuspid and pulmonary   2.1). Mean values are listed for selected weights along with
            flows. AO, Aorta; CaVC, caudal vena cava; LA, left atrium;   95% prediction intervals. However, these prediction inter-
            LV, left ventricle; PA, pulmonary artery; PV, pulmonary
            valve; RA, right atrium; RAu, right auricle; RC, NC, right   vals are quite wide, especially for larger dogs, and may
            and noncoronary cusps of aortic valve; RV, right ventricle;   encompass some degree of LV enlargement. Somatotype and
            RVO, right ventricular outflow tract. (Modified from Thomas   breed may have additional influence on normal echo values
            WP et al.: Recommendations for standards in transthoracic   in some dogs. For example, healthy Boxers can have increased
            2-dimensional echocardiography in the dog and cat, J Vet   LV wall thickness and smaller aortic dimensions relative to
            Intern Med 7:247, 1993.)                             nonBoxer dogs, although chamber dimensions are compa-
                                                                 rable. Slightly higher LV wall thickness and chamber dimen-
                                                                 sions have been observed in Greyhounds compared with
            beam placement for standard measurements and calcula-  other dogs of comparable weight. Endurance training also
            tions can be a limitation.                           affects measured parameters, reflecting the increased cardiac
                                                                 mass and volume associated with frequent and sustained
            M-Mode Views                                         strenuous exercise. Normal measurements in cats are more
            Standard M-mode views are obtained from the right    uniform but also are influenced by body size (Table 2.2).
            parasternal transducer position. The M-mode cursor is   Chamber volume and ejection fraction are better estimated
            positioned with 2-D guidance using the right paraster-  from optimized 2-D frames using the modified Simpsons’
            nal short-axis view; angling the transducer so that the LV   method rather than M-mode images because of the greater
            appears as round as possible helps ensure the ultrasound   potential for  inaccurate  geometric  assumptions  from one-
            beam is oriented perpendicular to the axis of the LV. Some   dimensional measurements (see  Suggested Readings for
            clinicians prefer using the long-axis view to obtain images   further  information).  The  right  parasternal  four-chamber
   46   47   48   49   50   51   52   53   54   55   56