Page 20 - Tlahuizcalli CB-30_Neat
P. 20

detected,  and  therefore  increase  the  statistics  of   meaning of this second peak is unclear, as it is not
          the experiment  in a  shorter operation time.  There    certain if there was something in the experimental
          would be complications however, of operating in         setup that could be responsible for it.
          an  open  space,  since  the  PMT  coupled  to  the
          scintillator needs to be connected to an H.V supply,     In addition to this, the left ends of the distributions
          and the output signal needs to be sent to the circuit   seem to have considerably more counts than the
          for its processing; not to mention the considerations   right ends, so it may have a correlation with how
          needed to be made regarding the weather.                small the mean time is. Also, just like in the mean
                                                                  lifetime  experiment,  there  was  a  noise  present
           Another  important  aspect  was  the  resolution  of   towards  the  beginning  of  the  spectrum,  so  we
          the  MCA  employed  to  register  the  decay  times.    needed  to  apply  a  large  enough  delay  to  the
          From the calibration, only time intervals up to 5       signals, in order to keep them away from that noise.
          could be measured. A consideration to be had for        These  effects  distorted  the  distributions,  making
          future reproductions of this experiment, is that the    them less gaussian, and therefore harder to fit. This
          resolution of the MCA can limit the accuracy of the     is what made the identification and subtraction of
          measurements.  In  this  case,  there  was  a  noise    the background necessary.
          present towards the beginning of the spectrum that
          had to be removed manually from the data before          One way to continue this project, would consist on
          it was used for the curve fit.                          performing  the  same  experiment  but  under
                                                                  different  conditions,  maybe  even  using  different
           One way to give continuity to this work, would be      detectors and electronics. The purpose would be
          to perform a similar experiment, detecting muons        to identify if some of the effects mentioned here,
          with  scintillators,  but  in  different  places  with  a   like  the  second  peak  on  the  right  end  of  the
          noticeable altitude difference. This would allow to     spectrum,  are  still  present.  This  could  help  to
          compare the amount of muons detected in each            determine their origin and nature.
          case,  and  the  difference  of  particles  detected
          would correspond to the ones that decayed when           Finally, it is important to highlight that, just like the
          going from one altitude to the other. Knowing the       mean  lifetime  experiment,  the  detectors  were
          velocity  of  the  particles,  it  would  be  possible  to   placed inside a laboratory on the ground level of a
          calculate  the  mean  lifetime  of  moving  muons.      building.  Moreover,  the  upper  scintillator  was
          These  results  could  be  compared  with  the  ones    located just under a lamp (see fig.(8)). Despite the
          presented here and used to study relativistic effects   lamp  remaining  off  during  the  entirety  of  the
          like time dilation.                                     experiment, there could be associated effects that
                                                                  we didn't consider.
          B. Muon’s speed
                                                                  V. Conclusions
                                                     7
                                       8
          The  value     = 2.400 × 10 ± 1.47 × 10   /  ,
          obtained for the mean speed of the muons using          The two values obtained for the mean lifetime of
          the  mean  time  intervals  for  the     = 50      and   the  muon,  with  different  PMT  operating  voltages,
                                            1
             = 100      separations,  is  physically  consistent   were  consistent  with  the  reported  value  in  the
           2
          with special relativity, as the muon, being a massive   Review of Particle Physics, within a 3   range. The
          particle, it cannot travel faster than the speed of     value calculated for the mean speed of the cosmic
          light, but because  it was generated  in the upper      ray muons, was consistent with special relativity.
          atmosphere by cosmic rays, their velocity must be
                         8
          close to 3 × 10   /  .                                   In  the  mean  lifetime  experiment,  a  background
                                                                  was present in the time distributions, so it had to be
                                                                  measured and considered to get a correct value
           Nevertheless,  there  were  some  noticeable
          problems  with  the  time  distributions.  Towards  the   of the mean lifetime. The origin of this background
                                                                  is still unclear.
          right end of both of the original spectrums, a small
          peak  can  be  distinguished.  Despite  being  small     There  were  effects  observed  in  the  time
          compared to the peak used to adjust the gaussian        distributions for the muon's speed experiment, that
          curve, it was always present in the measurements        could not be fully explained, like the presence of a
          for  all  the  separation  distances.  The  origin  or

              19
                            Tlahuizcalli ISSN: 2448-7260                    Año 10 Núm. 30 septiembre-diciembre 2024
   15   16   17   18   19   20   21   22   23   24   25