Page 170 - Feline diagnostic imaging
P. 170

10.7 Doppler  171



















                                                                  Figure 10.31  Imaging showing hand position to obtain left
                                                                  parasternal long axis view in cats with moderate to severe heart
               Figure 10.28  An image highlighting the left ventricular   disease.
               outflow tract (LVOT) as it approaches the aortic valve (Ao).
                                                                    In Doppler studies of the heart, the most accurate veloci-
                                                                  ties are recorded when the ultrasound beam is aligned paral-
                                                                  lel to flow, which can be a challenging achievement in feline
                                                                  patients. The ultrasound beam needs to be within 20° from a
                                                                  line parallel to flow so that peak velocities will not be under-
                                                                  estimated.  Ultrasound  machines  that  offer  only  color  and
                                                                  pulsed‐wave  Doppler  should  be  avoided  because  you  will
                                                                  only be able to document normal blood flow using either of
                                                                  these.  An  artifact  called  “aliasing”  occurs  with  color  and
                                                                  pulsed‐wave  Doppler  once  the  velocity  exceeds  a  value
                                                                  termed the Nyquist limit. It is important to purchase an ultra-
                                                                  sound  machine  that  can  perform  color,  pulsed‐wave,  and
                                                                  continuous‐wave Doppler in order to obtain a complete echo-
                                                                  cardiographic study and evaluate high pathologic velocities.

               Figure 10.29  By moving slightly more cranial to the view   10.7.1  Color Doppler Imaging
               pictured in Figure 10.23 and dropping the probe handle toward
               the table, this left ventricular long axis view is obtained,   Color  Doppler  imaging  is  useful  to  locate  areas  of  high
               depicting the right atrium (RA), tricuspid valves (TV), and right   velocity or turbulent blood. A common color map registers
               ventricle (RV).
                                                                  blood flow moving toward the transducer as red and away
                                                                  from the transducer as blue. A color bar representing the
                                                                  selected  color  map  appears  on  an  image  when  color
                                                                  Doppler is activated. Think of the top of the image as an
                                                                  observation platform for the direction of blood flow. When
                                                                  the  right  parasternal  short  axis  view  of  the  heart  at  the
                                                                  heart  base  is  viewed  from  the  observation  deck,  the  PV
                                                                  opens  downward  toward  the  pulmonary  arteries  so  the
                                                                  normal  blood  flow  will  be  blue  as  it  moves  away
                                                                  (Figures 10.32 and 10.33). When the valves snap closed, a
                                                                  small  red  jet  which  looks  like  a  candle  flame  moving
                                                                  toward the platform can cause concern for a small amount
                                                                  of insufficiency. However, if this red jet travels less than
                                                                  1 cm from the valve, this is consistent with a physiologic
                                                                  process, not a pathologic process [4]. At this same level, be
               Figure 10.30  Image showing hand position to obtain left
               parasternal long axis view in cats with heart disease or sternal   careful not to overinterpret different colors adjacent to the
               contact due to age.                                aorta.  As  the  blood  moves  from  the  TV  to  the  right
   165   166   167   168   169   170   171   172   173   174   175