Page 173 - Feline diagnostic imaging
P. 173

174  10  Normal Cardiovascular Imaging
            the PV. As the PVs open and close, the spectral Doppler   valve closes and before it opens again and be depicted below
            image  will  be  a  waveform  with  laminar  flow  depicted   the baseline on the spectral Doppler imaging.
            below the baseline (Figure 10.37). A sharp line can be seen
            when the valve opens and closes; however, this does not   10.7.6  Tricuspid Valve
            depict insufficiency. The normal laminar flow at the PV in
            cats is typically less than 1.2 m/s [12]. Insufficiency would   Spectral Doppler of the TV appears similar to the MV with
            be above the baseline after the valve closes and before it   a decreased amplitude in the normal flow (Figure 10.40).
            opens again. Insufficiency at this valve greater than 1.9 m/s
            is  consistent  with  pulmonary  hypertension.  If  the  color   10.7.7  Aortic Valve
            Doppler is insensitive on your machine, then mapping of   The aortic valve can be the most technically difficult valve to
            the  valve  by  using  pulsed‐wave  or  continuous‐wave   align properly for Doppler evaluation. With the footprint of
            Doppler imaging in the right ventricular outflow tract will   the probe in the same location, the angle of the probe may
            isolate any insufficiency. Mapping means moving the box   need to be steepened by pulling the handle toward the table
            or cursor above the valve in the outflow tract, keeping in   and caudally toward the body wall. The normal velocity at
            mind theat insufficiency does not always go directly above   the aortic valve is typically less than 1.2 m/s [12]. The velocity
            the valves.
                                                              at the level of the aortic valve (Figure 10.41) and within the
                                                              LVOT (Figure 10.42) should be documented. If narrowing is
                                                              present in the LVOT due to hypertrophy of the IVS, continu-
            10.7.5  Mitral Valve
                                                              ous‐wave Doppler should be used because higher velocities
            On the left parasternal long axis view of the heart, the MV   cannot  be  recorded  with  pulsed‐wave  Doppler.  Increased
            and aortic valve are normally seen adjacent to each other,   velocity in the LVOT is consistent with outflow obstruction
            assisting in identification of left versus right heart. From the   and can occur in older cats with asymmetric hypertrophic
            observation  deck,  the  blood  flow  moves  from  the  LA   cardiomyopathy or in a young cat with mitral dysplasia.
            through the MV toward the apex of the LV. The motion of
            the MV is normally biphasic but in cats, the pulsed‐wave
            Doppler appears as one peak due to their elevated heart rate   Acknowledgment
            (Figures 10.38 and 10.39). The normal velocity of flow at the
            MV is usually less than 1.0 m/s. Sharp lines can be seen at   Grateful thanks to Dr Pete Velotas and his wonderful cat
            the opening and closing of the MV leaflets, which do not   Ody for providing a normal volunteer for acquiring these
            indicate insufficiency. Insufficiency would occur after the   echocardiographic images.































            Figure 10.38  Pulsed-wave Doppler image of the mitral valve shows normal biphasic waveforms above the baseline. Notice the
            single-phase waveform (arroe); this depicts a normal finding when the heart rate is elevated.
   168   169   170   171   172   173   174   175   176   177   178