Page 324 - Zoo Animal Learning and Training
P. 324

308    Tasks for the Veterinary Assistant


          Checking for Light Leaks                           headache, so protecting yourself is very important. Film
                                                             processor rollers are what pulls the radiography film
          Door seals tend to wear out or get torn loose just from   through the processor. These should be wiped off daily
          daily wear and tear. To check the darkroom for leaks,   to weekly depending on the number of radiographs
          close the door, turn off the light, and note the presence   taken at the clinic. The least often the more they need to
          of any daylight that might be entering around the door   be cleaned as the chemicals tend to dry on the rollers
          or pass‐through box. If there is any light from cracks   causing artifacts on the  finished film.
          around the door or any other source, notify the office   Professional processor maintenance  is important  to
          manager.                                           maintain the life of a processor. Read the manufacturer’s
            Check the safety light as this can deteriorate as the   recommendations and discuss this with the office
          bulb is worn. Close the door, turn off the overhead light,     manager to set up the service.
          and turn the safety light on. Place an unexposed film on   Floors and counters can become wet. All spills or wet
          the counter with a metal object such as a paper clip on   areas need to be cleaned up and dried immediately. Wet
          the film. Let it set for 2 minutes then develop the film.   spots will make artifacts on the film. Keep paper towels
          If the object appears on the film, the safety light is not   stocked in the room for wiping up spills and cleaning the
          working properly or there is some other source of light   counter. Mop the floor whenever it becomes wet or at
          in the room.                                       least weekly. Keep a box of examination gloves, an apron,
            Check to see if the flasher blocker is working prop-  and goggles in the room to be used whenever working
          erly. If the light doesn’t work, it can be as simple as   with chemicals. Maintain an adequate supply of materials
          changing the bulb. Make sure to get the proper watt-  and chemicals in inventory. Note the quantity of each
          age bulb, as too bright will cause an artifact on the film   whenever using these items. Add needed items to the
          and too dim will not show the label very well. Always   want list or notify the inventory manager as supplies
          follow the manufacturer’s directions for product   become low. Allow for sufficient time for ordering and
          maintenance.                                       delivery.
            Ideally, there are two areas to a darkroom: a wet side   The veterinary assistant can develop daily, weekly, and
          and a dry side. The processor side is considered the wet   monthly maintenance charts for the darkroom and mate-
          side and the countertops as the dry side. There needs to   rials stored in it. Compose a chart of rows and  columns to
          be sufficient dry counter space for working with cassettes   meet the needs and policies of your practice.
          and film boxes. If a flasher block is used for film   Maintenance tasks include the following.
          identification, make sure there is enough room to   1.  Counters should be wiped down daily with dilute
          accommodate the film in the flasher without having to   disinfectant. Paper towels are restocked, and the
          bend the film. Even a slight bend in an unprocessed film   disinfectant is replenished as needed.
          can cause an artifact. The cassettes should be able to lie   2.  Developer and fixer chemical levels are checked
          flat on the counter, with the lids open so they don’t acci-  daily and replenished as needed.
          dentally fall.                                     3.  Note items that need to be added to a want list or
                                                                need repair. Next time you are near the want list,
          Processor Maintenance                                 copy the information down or pass the note on to
                                                                the inventory manager if more appropriate.
                                                             4.  File radiographs daily.
          The film processor with the developer and fixer tanks   5.  Floors are mopped weekly.
          are on the wet side of the room. The processor needs   6.  View boxes are cleaned weekly and bulbs replaced
          weekly attention. The developer and fixer tanks need to   as needed. Any glass‐cleaning liquid will do. Dry
          be refreshed or changed as per the quality of the radio-  thoroughly so no streaks are left. View boxes may be
          graphs.  The developer and fixer chemicals  are pulled
          from tanks as the film moves through the machine.
          These tanks usually located on the floor of the darkroom
          under the processor. When these get too low for the
          hoses to pump fresh chemicals into the machine or they   Learning Exercise
          have reached the expiration date, they need to be
          refreshed or changed. Follow the instructions on the   Take a look at the automatic processor in your
          outside of the developer and fixer chemical concen-  school or clinic if there is one. Lift the covers and
          trates for refreshing or changing. PPE required for this   look at the rollers. Are they clean? Do they need
          task are goggles, waterproof apron, and gloves. It seems   attention? Are the developer and fixer tanks both
          that no matter how careful you are there is always spill-  sufficiently filled? Is the developer tank dark brown
          age of water or  chemicals. These chemicals are not   or light brown?
          highly toxic but  cause skin drying and for some a
   319   320   321   322   323   324   325   326   327   328   329