MSI 6260

1. Turn off scale, open case and close S2-8
2. Close door and turn scale on, do not press buttons
3. Remove access plug to reveal 7 segment display and cal pot and pushbutton
4. Press the Power button for a long second
5. Press and release the Cal button
6. Turn cal pot until 2 is displayed
7. Press the cal pb
8. At the “A” “n” prompt, press the cal pb
9. “0” will flash, remove weight, when stabile, press the cal pb.
10. Unit flashes “0”, “C”, Use cal pot to adjust close to zero and press the cal pb
11. Unit flashes “0”, “F”, Use cal pot to adjust exactly to zero and press cal pb
12. On the indicator unit, press the ZERO button
13. Display will flash “L”, lift weights and press CAL pb
14. “L”, “C” flashes, use cal pot to adjust weight reading close to test weight value
15. Press Cal PB, “L”,”F” flashes, use cal pot to adjust weight reading to exactly match test

weight value.
16. Press the Cal PB, unit should flash “U”, unload unit and press cal PB
17. Unit will flash “r”, Open cal switch in indicator and turn cal pot to 8 to exit.

NCI

NCI 61XX
1. Press SW-1
2. Press ZERO until CAL is displayed
3. Press TEST to access cal menu
4. Press TEST
5. Press TEST, Load 0 is displayed
6. Press TEST to set zero calibration
7. “LOAD XX” is displayed
8. Press ZERO to toggle between cal weight choices
9. Load scale and press TEST key
10. “DONE” is displayed, unload scale and press TEST.

NCI 6700

1. Move SW 1 to menu mode Press ZERO key twice
2. “CAL” is displayed
3. Press TEST 4 times until “LOAD 0” is displayed
4. Clear all weight and press the TEST key
5. Place cal weight specified by XX and press the TEST key
6. Diag is displayed, press the TEST key.
7. Move SW1