NCI 7050

1. Press the ESCAPE key, display will prompt “Clear data? Yes/no”
2. Press the ESCAPE key and hold for until 5 softkeys appear
3. Press SETUP softkey and ENTER
4. Press CALIBRATION softkey and key in 999 and ENTER.
5. Follow prompts to calibrate local and remote bases. Then Select EXIT

NCI 7620
1. Move SW1 to Menu Mode
2. “CAL” is displayed
3. Press UNITS key until “LOAD 0” is displayed
4. Clear weight and press UNITS key
5. LOAD XXX is displayed, use the ZERO key to toggle between weight choices
6. Place cal weight on scale and press UNITS key
7. “done” then “CAL” is displayed, move SW1

NCI 7800

1. Move SW1 to Menu Mode
2. Press ZERO Key until CAL is displayed
3. Press TEST key until “LOAD 0” is displayed
4. Clear weight and press TEST key
5. “LOAD XX” is displayed, Load test weight and press TEST key
6. “DIAG” is displayed, slide SW1

NCR

NCR 7870

1. Raise the top plate and press the cal switch, lower plate quickly
2. Press ZERO
3. Add weight to scale followed by pressing the ZERO key until finished.

NCR 7872/7875/7880

1. Press Cal button under platter
2. Follow prompts, placing 5, 10, 15, and 30 lbs on the scale