Suguna Management System       Ver 1.0 / SOP / FQL / P2 - 32 Page  1 of 1



























           32. Estimation of Sodium & Potassium


                                                                   accurately  0.2542gm  of  sodium  chloride  &
           SCOPE
                                                                   0.1909gm of potassium chloride and make up 1lit
           To estimate the sodium & potassium in feed & feed       distilled water in std.volumetric flask. This solution
           stuffs.
                                                                   containing 100 ppm of sodium & potassium.

           PRINCIPLE
                                                                PROCEDURE
           The solution to be analyzed is discharged through an
           atomizer in a fine mist into a chamber, when it is drawn   FSwitch on the flame photometer, then switch on air
           into a flame. By an optical system the light produced by   compressor  and  maintain  10  PSI  with  adjust  air
           the  combustion  of  the  elements  in  the  vaporized   valve in the instrument.
           solution  is  conducted  through  appropriate  filters  to   FThen open LPG gas cylinder valve after open gas
           impinge upon a photoelectric cell which activates a
           galvanometer. Under proper operative conditions the     valve in the instrument, fire the flame and should
           concentration of sodium or potassium in the solution    adjust the blue flame sharp in top of the flame.
           can be estimated from the reading of the galvanometer.   FTake distilled water in 50ml beaker put suction tube
           The  instrument  is  equipped  with  an  amplifier  which   in distilled water beaker. Set the reading zero with
           permits the analysis of solutions with greatly varying   use zero set knob.
           concentrations of sodium and potassium.
                                                                FThen  take  standard  solution  in  another  25ml
                                                                   beaker, after removing the distilled water beaker put
           APPARATUS
           FFlame Photometer                                       the suction tube in standard solution beaker and
           FLGP Gas cylinder                                       set  100  in  sodium  level  and  also  set  100  in
           FBeaker 50ml, 25ml                                      potassium level with use set knob. After removing
                                                                   the  std.solution,  recheck  the  zero  reading  with
           FCrucible
                                                                   distilled water.
           FVolumetric flask 100ml
                                                                FTake sample solution (the solution ash obtained in
                                                                   previous  experiment  is  used  for  estimation  of
           CHEMICALS
           FStandard  Sodium  &  Potassium  Solution:  Weigh       calcium and phosphorus.



          CALCULATION
           % of Sodium = Digital Reading of Sodium / 200

           % of Potassium =  Digital Reading of Potassium / 200



          Prepared by :                                                             Approved by :