Page 504 - Med Plaza and Cancer Center
P. 504

4/TEST PROCEDURES


              4.1 Testing the Switch Gauge Circuitry            7.  Return each low contact assembly to its original
                                                                   position (mark placed on transparent tape).
              The following procedure can be used to check switch
              gauge circuitry. See Figures 15 and 16.           8.  Repeat steps 5 through 7 to test each high pressure
                                                                   contact.
              Tools Required:
              #2 Phillips screwdriver                           9.  Replace the louvered panels.
              1/16 hex wrench
              Transparent tape                                  Note:  This test only checks switch gauge electronics; it
                                                                does not verify alarm set points.
              1.  Remove the left and right louvered panels by
                 removing the Phillips head screws located on the top
                 of each panel. Figure 15.

              2.  Verify that the readings indicated on the switch gauge
                 and the associated line pressure gauge match to
                 within:
                  6 psig for the nominal 50 psi gases
                  18 psig for the 180 psi gases
                  2 inches mercury for the vacuum and evacuation

              3.  Place a suitable length of transparent tape across the
                 face of each switch gauge. The tape should be long
                 enough so that the ends of the tape clear the front
                 panel. Figure 16.
              4. Mark the exact location of the contact assembly’s
                 brass pins on the transparent tape.
              5.  Using the 1/16 hex wrench, individually turn each low
                 contact so it just touches the gauge needle.

              6.  The alarm must sound on contact. If no alarm sounds
                 the panel needs repair.                        Note:  Lip on louvered panel not shown. When replacing,
                                                                make sure lip is properly positioned.
              Note: If remote alarms or monitors are connected, these
              devices will be activated.
                                                                                                       Figure 15
                                                                                          Louvered Panel Removal





























                                                                                                       Figure 16
              Note:  Shorting bar has been omitted for clarity.                    Testing the Switch Gauge Circuitry


              0178-0135-000  Rev. A                                                                         4-1
   499   500   501   502   503   504   505   506   507   508   509