Page 505 - Med Plaza and Cancer Center
P. 505

4/TEST PROCEDURES


              4.2 Setting Alarm Trigger Points                  2. Place a suitable length of transparent tape across the
                                                                   face of the switch gauge. The tape should be long
              The alarm trigger points for each gas can be set on an  enough so that the ends of the tape clear the front
              individual basis by using either of the following    panel. Figure 17.
              procedures:
                                                                3. Mark a straight line on the tape between the desired
              Note:  The accuracy of this first “simplified” procedure is  low pressure alarm setting on the scale and the pivot
              limited. The second procedure which requires the use of  (point) of the switch gauge needle.
              a test apparatus is more accurate and is recommended
              by BeaconMedæs.                                   4. Using the 1/16 hex wrench, individually turn the low
                                                                   (left) contact to the desired set point by aligning the
              Note:  In accordance with NFPA Standard 56F (3-3), set  center of the brass contact with the edge of the line
              the alarms, “to indicate if the pressure in the piping  on the tape.
              system increases or decreases 20 percent from the
              normal operating pressure”.                       5. Repeat steps 3 through 4 to set the high (right)
                                                                   pressure contact. Repeat this procedure for each
              Procedure #1                                         additional gas module.
              Tools required:
              Phillips screwdriver                              6. Remove tape and replace the louvered panel.
              1/16 hex wrench
              Transparent tape                                  Procedure #2
                                                                Figure 18
              1.  Remove the left and right louvered panels by
                 removing the Phillips head screws located at the top  The alarm trigger points for each gas (and vacuum) can
                 of each panel. Figure 15.                      be tested and set on an individual basis by using the
                                                                following procedure. This procedure requires a
                                                                special test apparatus as detailed.











































                                                                                                       Figure 17
              Note:  Shorting bar has been omitted for clarity.              Alarm Trigger Adjustment (Procedure #1)



              4-2                                                                              0178-0135-000  Rev. A
   500   501   502   503   504   505   506   507   508   509   510