Operation
Set-Up Procedure Multi-Signal Alarm Module Circuit Board
Multi-Signal Module
SETUP
Changing signal options: button
1. Loosen two screws on front of alarm
panel. Hinge open front cover.
2. Locate SETUP button on circuit board
of multi-signal module to be adjusted.
3. Press and release SETUP button. Red LED Green LED
Audible alarm will chirp and both
SIGNAL 1
green and red indicators for top
location (Signal 1) will illuminate. SIGNAL 2
4. Using either or buttons, enable SIGNAL 3
or disable Signal 1 location. When red SIGNAL 4
indicator is illuminated, location is
SIGNAL 5
ENABLED. Turning off red indicator
disables location. Signal locations can
not be disabled if wired to a closed-
contact circuit.
5. Press and release TEST button on
front of alarm panel. Both green and
red indicators for second location
(Signal 2) will be illuminated.
6. Using either or buttons, enable TEST
or disable Signal 2 location. When red button
LED is illuminated, location is enabled.
Turning off red LED disables location.
7. Repeat steps 5 and 6 above for
remaining signal locations (Signal 3, 4,
and 5).
8. Press and release TEST button to
save changes. LEDs will sweep from Up and down
top to bottom to indicate end of set-up
buttons
sequence has been reached. Multi-
signal module will return to monitoring
NOTE:
mode.
Audible alarm will chirp with each button push.
If module is idle for more than one minute during
set-up procedure, module will chirp three times,
return to monitor mode and will default to previ-
ous settings. No changes to settings will be
saved.
The set-up must be completed in order for
changes to be saved.
38
Part No. 6-847684-00 Rev. E00