Operation



              Set-Up Procedure                                           Multi-Signal Alarm Module Circuit Board


             Multi-Signal Module
                                                                 SETUP
             Changing signal options:                             button
             1. Loosen two screws on front of alarm
                panel.  Hinge open front cover.
             2. Locate SETUP button on circuit board
                of multi-signal module to be adjusted.
             3. Press and release SETUP button.                   Red LED           Green LED
                Audible alarm will chirp and both
                                                                                   SIGNAL 1
                green and red indicators for top
                location (Signal 1) will illuminate.                                SIGNAL 2
             4. Using either  or  buttons, enable                                 SIGNAL 3
                or disable Signal 1 location.  When red                             SIGNAL 4
                indicator is illuminated, location is
                                                                                    SIGNAL 5
                ENABLED. Turning off red indicator
                disables location.  Signal locations can
                not be disabled if wired to a closed-
                contact circuit.
             5. Press and release TEST button on
                front of alarm panel.  Both green and
                red indicators for second location
                (Signal 2) will be illuminated.
             6. Using either  or  buttons, enable               TEST
                or disable Signal 2 location.  When red           button
                LED is illuminated, location is enabled.
                Turning off red LED disables location.
             7. Repeat steps 5 and 6 above for
                remaining signal locations (Signal 3, 4,
                and 5).
             8. Press and release TEST button to
                save changes.  LEDs will sweep from                                      Up and down
                top to bottom to indicate end of set-up
                                                                                            buttons
                sequence has been reached.  Multi-
                signal module will return to monitoring
                                                               NOTE:
                mode.
                                                               Audible alarm will chirp with each button push.
                                                               If module is idle for more than one minute during
                                                               set-up procedure, module will chirp three times,
                                                               return to monitor mode and will default to previ-
                                                               ous settings.  No changes to settings will be
                                                               saved.

                                                               The set-up must be completed in order for
                                                               changes to be saved.