Operation
Start-Up and Checking
1. Turn on electrical power to alarm
panel.
2. Make following observations:
• POWER ON light illuminates on Status LED
Indicators
front of alarm panel.
• Digital display and multi-signal
modules warm-up for ten seconds.
During warm-up, no audible or LED Display
visual alarms activated. After
warm-up, audible and visual
alarms are initiated for any active
alarm.
3. Press MUTE button to silence
audible alarm. Digital Display Module
4. Pressurize piping system (medical gas
and vacuum).
5. Make following observations:
• Digital display modules display
actual pipeline pressures or
vacuum levels.
• If digital display module pipeline SIGNAL #1
pressures and vacuum levels are
within pre-set limits, NORMAL SIGNAL #2
indicator is illuminated. Status LED
• If digital display module pipeline Indicators SIGNAL #3
pressures and vacuum levels are
SIGNAL #4
outside pre-set limits, status
indicators show HIGH or LOW.
SIGNAL #5
• If a fault code is displayed (-- FF 11 --,
- - FF 22 --, -- FF 33 -- or -- FF 44 --) refer to Multi-Signal Module
Troubleshooting Guide.
• Multi-signal modules indicate
current status from dry-contact
switches. Unused signals can be
disabled. Refer to Set-Up
Procedure.
NOTE:
If any module appears to be malfunctioning, refer
to Troubleshooting Guide.